68 Agilent 7890B Installation
1 Installing the GC
Step 20. When the system stabilizes, run one injection
Perform the checkout procedure as described in
"Chromatographic Checkout" in the Operation Manual.
1 Enter the parameters for the checkout procedure.
If using an Agilent data system, use it to create a checkout
method.
If not using a data system, enter the setpoints using the
GC keyboard.
See the following topics in the Operation Manual for the
parameters for your detector:
FID checkout
FPD
+
checkout
FPD
+
checkout (Japan)
NPD checkout
TCD checkout
μECD checkout
2 When the GC becomes Ready (Not ready light turns off), make
the injection and start the run.
For an ALS injection, press [Start] on the GC or in the data
system, as appropriate.
For a manual injection, inject the sample and press [Start].