4 Operating the 850-DS
850-DS Operator’s Manual 39
Option Function
Name 30-character field for the name of this method.
Vessel Temp Desired vessel temperature for this method.
Bath Temp Desired bath temperature for this method.
Spindle Speed Desired spindle speed for this method.
Media Volume Volume of media initially in the vessels.
Rotation Start Delay Time after the start of the method before the spindle rotation begins.
Profile Print Interval Specifies the frequency that dissolution apparatus operational parameters are
printed on the optional printer.
Apparatus 1/2/5/6 Select the appropriate apparatus type.
Initial Temp Takes a temperature reading of the vessels prior to dropping a dosage at the
start of a method.
Initial Sample Takes a sample prior to dropping a dosage at the start of a method.
Final Temp Takes a temperature reading of the vessels at the end of the method.
Enable Final Spin After the final timepoint, this option runs the spindles at a high speed in order to
try to get all remaining drug into suspension.
RPM Rotational speed used for the Final Spin.
Duration Duration the final spin will be executed.
Clear Erases all values entered and returns them to default values.
Import Allows user to import method from SD card.
Cancel Exits the method entry screens and returns to the Main screen without saving
any parameters.
Save Saves all method parameters and returns to the Main screen.