5-4
Service
Recommended Test Equipment
Frequency
Counter
Resolution: 0.1 Hz 5316B Internal Reference Oscillator
(Manual Adjustment)
(A)
Function
Generator
Freq. range: 1 kHz
Amplitude: 4 Vpk
DC output: ±4 Vdc
33120A AM Level and Distortion
AM Level
AM Level:FE
AM Modulator
Motherboard Audio Path
(A)
(A)
(A)
(A)
(A)
Measuring
Receiver
FM Accuracy: ±2% of reading ±1 digit
AM accuracy: ±2% of reading ±1 digit
Range: 250 kHz to 1000 MHz
Filters: 300 Hz high-pass; 15 kHz
low-pass
Detectors: Peak+
8902A FM Accuracy
FM Distortion
AM Accuracy
Phase Modulation Distortion
Residual FM
RF Level Accuracy
AM Modulator
Motherboard Audio Path
(P)
(P)
(P)
(P)
(P)
(P)
(A)
(A)
Oscilloscope Bandwidth: 1 GHz 54100A Pulse Modulation On/Off Ratio
Pulse Modulation Rise Time
(P)
(P)
Oscilloscope Bandwidth: 100 MHz 54600B Internal Timebase: Aging Rate (P)
Power Meter Instrumentation accuracy: ±0.5%
Power reference accuracy: 0.9%
438A RF Level Accuracy
Power Level Accuracy
Detector Offset
Output Level
Predistortion& Detector Offset
Prelevel
Output Level:FE
HF Power Level Accuracy
(P)
(P)
(A)
(A)
(A)
(A)
(A)
(A)
Power Sensor Freq. range: 100 kHz to 4.2 GHz
Power range: 30 dBm to +13 dBm
Maximum SWR:
100 kHz to 300 kHz 1:1.6
300 kHz to 1 MHz 1:1.2
1 MHz to 2 GHz 1:1.1
2 GHz to 4.2 GHz 1:1.3
Cal factor accuracy (RSS): 1.6%
8482A RF Level Accuracy
Power Level Accuracy
Detector Offset
Output Level
Predistortion & Detector Offset
Prelevel
Output Level: FE
HF Power Level Accuracy
(P)
(P)
(A)
(A)
(A)
(A)
(A)
(A)
Power Sensor
(Low)
Freq. range: 100 kHz to 4.2 GHz
Power range: 70 dBm to 20 dBm
Maximum SWR:
100 kHz to 300 kHz 1:1.2
300 kHz to 2 GHz 1:1.15
2 GHz to 4.2 GHz 1:1.4
Power linearity (30 to 20 dBm): ±1%
Cal factor accuracy: < 1.6%
8481D
Option H70
RF Level Accuracy (P)
Primary
Frequency
Standard
Frequency: 10 MHz
Stability: > 1 x 10
10
/year
5071A or
5061A/B
Internal Timebase: Aging Rate
Internal Reference Oscillator
(Manual Adjustment)
Time Base DAC
(Automated Adjustment)
(P)
(A)
(A)
Table 5-1. Recommended Test Equipment
Instrument Critical Specifications Recommended
Model
Use
Performance Test - (P)
Adjustment - (A)
Supp. Verification Test - (V)

Table of Contents

Other manuals for Agilent Technologies 8648A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8648A and is the answer not in the manual?

Agilent Technologies 8648A Specifications

General IconGeneral
BrandAgilent Technologies
Model8648A
CategoryTest Equipment
LanguageEnglish

Related product manuals