7-44
Adjustments
Agilent Service Support Software
9. The New Test Equipment window is displayed:
Enter the following parameters and then click OK:
Serial Number of the new equipment (8).
GPIB Address of the new equipment (9). This address must be in the range of 0
through 31 and it should not conflict with any other instrument
address already present in the test setup. Note: The power sensor
must be assigned GPIB address 1 (negative 1).
Cal Due Date the projected calibration due date of the new equipment (10).
Trace Number the calibration tracking number (11). This is the last required item.
Calibration Type (optional) is used only when special calibration data is associated
with the device being added, and only as a function setup by the
factory for equipment requiring specific calibration data (14). The
only devices currently requiring this feature are power sensors.