8-119
Performance Tests
8648C Test Record
RF Level Accuracy performance test is not required.
NOTE If the automated Power Level Accuracy performance test is performed, the
RF Level Accuracy performance test is not required.
Table 8-64. RF Level Accuracy Performance Test - Part 3
Frequency
(MHz)
Amplitude
(dBm)
Limits (dBm)
Uncertainty
(±dB)Lower Measured Upper
3.0 13 11.5 14.5 0.23
249.9 18 16.5 19.5 0.23
400 18 16.5 19.5 0.23
999.9 18 16.5 19.5 0.23
1500 17 15.5 18.5 0.23
2100 15 13.5 16.5 0.40
2500 13 11.5 14.5 0.58
3200 11 9.0 13.0 0.58
Table 8-65. RF Level Accuracy Performance Test with Option 1EA and 1E6 -
Part 3
Frequency
(MHz)
Amplitude
(dBm)
Limits (dBm)
Uncertainty
(±dB)Lower Measured Upper
3.0 13 11.5 14.5 0.23
249.9 18 16.5 19.5 0.23
400 18 16.5 19.5 0.23
999.9 18 16.5 19.5 0.23
1500 17 15.5 18.5 0.23
2100 15 13.5 16.5 0.40
2500 13 11.5 14.5 0.58
3200 11 9.0 13.0 0.58