EasyManua.ls Logo

Agilent Technologies 8648A - Page 472

Agilent Technologies 8648A
481 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9-6
Supplemental Verification Tests
CW Frequency Accuracy Supplemental Verification Test
Table 9-5. 8648A Option 1E5 Frequency Accuracy Supplemental Verification Test
Frequency
(MHz)
Limits (Hz)
Uncertainty
(±Hz)Lower Measured Upper
0.5 1.0 1.0 1.0
1 1.0 1.0 1.0
10 2.0 2.0 1.0
50 8.0 8.0 1.0
100 15.0 15.0 1.0
500 75.0 75.0 1.0
1000 150 150 1.0
Table 9-6. 8648B Option 1E5 Frequency Accuracy Supplemental Verification Test
Frequency
(MHz)
Limits (Hz)
Uncertainty
(±Hz)Lower Measured Upper
0.5 1.0 1.0 1.0
1 1.0 1.0 1.0
10 2.0 2.0 1.0
50 8.0 8.0 1.0
100 15.0 15.0 1.0
500 75.0 75.0 1.0
1000 150 150 1.0
2000 300 300 1.0

Table of Contents

Other manuals for Agilent Technologies 8648A

Related product manuals