9-7
Supplemental Verification Tests
CW Frequency Accuracy Supplemental Verification Test
Table 9-7. 8648C Option 1E5 Frequency Accuracy Supplemental Verification Test
Frequency
(MHz)
Limits (Hz)
Uncertainty
(±Hz)Lower Measured Upper
0.5 1.0 1.0 1.0
1 1.0 1.0 1.0
10 2.0 2.0 1.0
50 8.0 8.0 1.0
100 15.0 15.0 1.0
500 75.0 75.0 1.0
1000 150 150 1.0
2000 300 300 1.0
3000 450 450 1.0
3200 480 480 1.0
Table 9-8. 8648D Option 1E5 Frequency Accuracy Supplemental Verification Test
Frequency
(MHz)
Limits (Hz)
Uncertainty
(±Hz)Lower Measured Upper
0.5 1.0 1.0 1.0
1 1.0 1.0 1.0
10 2.0 2.0 1.0
50 8.0 8.0 1.0
100 15.0 15.0 1.0
500 75.0 75.0 1.0
1000 150 150 1.0
2000 300 300 1.0
3000 450 450 1.0
3200 480 480 1.0
4000 600 600 1.0

Table of Contents