8- 7
Safety and Regulatory Information
Safety Considerations
General
WARNING To prevent electrical shock, disconnect the analyzer from mains before cleaning. Use a
dry cloth or one slightly dampened with water to clean the external case parts. Do not
attempt to clean internally.
WARNING If this product is not used as specified, the protection provided by the equipment could be
impaired. This product must be used in a normal condition (in which all means for
protection are intact) only.
CAUTION This product is designed for use in Installation Category II and Pollution Degree 2 per IEC
1010 and 664 respectively.
CAUTION VENTILATION REQUIREMENTS: When installing the product in a cabinet, the convection
into and out of the product must not be restricted. The ambient temperature (outside the
cabinet) must be less than the maximum operating temperature of the product by 4 C for
every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is
greater that 800 watts, then forced convection must be used.
WARNING Install the instrument according to the enclosure protection provided. This instrument
does not protect against the ingress of water. This instrument protects against finger
access to hazardous parts within the enclosure.