1-78
Making Measurements
Using Limit Lines to Test a Device
1. To access the limits menu and activate the limit lines, press:
2. To establish the start frequency and limits for a sloping limit line that tests the low side of the filter,
press:
3. To terminate the lines and create a sloping limit line, press:
4. To establish the start frequency and limits for a sloping limit line that tests the high side of the filter,
press:
5. To terminate the lines and create a sloping limit line, press:
You could use this type of limit to test the shape factor of a filter.
System
LIMIT MENU
LIMIT LINE
LIMIT LINE ON
EDIT LIMIT LINE
CLEAR LIST
YES
ADD
STIMULUS VALUE
123
M/
UPPER LIMIT
65
x1
LOWER LIMIT
200
x1
DONE
LIMIT TYPE
SLOPING LINE
RETURN
ADD
STIMULUS VALUE
125
M/
UPPER LIMIT
26
x1
LOWER LIMIT
200
x1
DONE
LIMIT TYPE
SINGLE POINT
RETURN
ADD
STIMULUS VALUE
144
M/
UPPER LIMIT
26
x1
LOWER LIMIT
200
x1
DONE
LIMIT TYPE
SLOPING LINE
RETURN
ADD
STIMULUS VALUE
146
M/
UPPER LIMIT
65
x1
LOWER LIMIT
200
x1
DONE
LIMIT TYPE
SINGLE POINT
RETURN

Table of Contents

Other manuals for Agilent Technologies 8720ES

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8720ES and is the answer not in the manual?

Agilent Technologies 8720ES Specifications

General IconGeneral
BrandAgilent Technologies
Model8720ES
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals