Chapter 1 25
Overview
Service Equipment You Will Need
Microwave Source 1
(for E4446A, 47A, 48A)
Frequency: 10 MHz to 50 GHz
Frequency Resolution: 1 Hz
Harmonic level: < -30 dBc
Amplitude range: -20dBm to +13
Amplitude resolution: 0.02
VSWR: < 20 GHz: 1.6:1
≤ 40 GHz: 1.8:1
≤ 50 GHz: 2.0:1
E8257D
b 83650A/B
(Option 001, 008)
PSG
cd
A, P, T
Microwave Source 2
(only required for Third
Order Intermodulation and
Gain Compression)
Frequency: 10 MHz to 26.5 GHz
Frequency Resolution: 1 Hz
Harmonic level: < -30 dBc
Amplitude range: -20dBm to +13
Amplitude resolution: 0.02
VSWR: < 20 GHz: 1.6:1
≤ 31 GHz: 1.8:1
PSG
c 83630A/B
83640A/B
83650A/B
(Option 001, 008)
(One sweeper
requires Option
001 for PSA
Option 110
testing)
A, P, T
Function Generator 1 Frequency: 10 Hz to 300 kHz
Amplitude Resolution: 0.1 mv
Harmonic Distortion: -35 dBc
33250A 33120A
33120A
(Option 001)
A, P
Audio Source 1
(for Option 107)
THD: 20 Hz to 125 kHz < -68 dB Stanford
Research DS360
Stanford
Research
Systems,
Sunnyvale, CA
P
Counters
Universal Counter Frequency: 10 MHz
Gate time: 10 to 100 seconds
Must be capable of measuring signal at
+7 dBm (0.5 Vrms)
53132A 53131A P, T
Meters
Digital Multimeter AC Accuracy: ± 0.31% of reading 3458A A, T
Power Meter Dual Channel
Absolute Accuracy: ± 0.5%
Resolution: 0.01 dB
Power Reference Accuracy: 1.2% (± 0.9% rss)
Compatible with 8480 series power sensors
dB relative mode
E4419B E4419A
N1912A
A, P
Table 1-3 Required Test Equipment for PSA Series (Continued)
Instrument Critical Specifications Recommended
Agilent Model
Number
Alternative
Agilent Model
Number
Use