406 Chapter 11
Assembly Replacement Procedures
Front Frame Subassemblies
Replacement
1. Place the keypad into position in the bezel. Press on the keypad to
engage the alignment pins and to seat the keys.
2. Place the keyboard over the bezel alignment pins and carefully slide
to lock into place.
3. Refer to Figure 11-63. Place the keyboard/bezel assembly face down
on top of the dress panel and subpanel on a flat surface. Install the
front frame over the keyboard/bezel assembly. Using the T-10 driver,
replace the 10 screws (1) that secure the bezel and keypad to the
front frame. Torque to 9 inch pounds.
4. Press the RPG knob and volume knob onto the control shafts.
5. Replace the display/rubber mount assembly. Refer to the last section
of the “Display Replacement” procedure.