106 Agilent 1260 Infinity RID User Manual
7 Error Information
General Error Messages
Cover Violation
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the
foam is removed while the lamps are on (or if an attempt is made to switch on
for example the lamps with the foam removed), the lamps are switched off,
and the error message is generated.
Probable cause Suggested actions
1 The top foam was removed during operation.
Please contact your Agilent service
representative.
2 Foam not activating the sensor.
Please contact your Agilent service
representative.