EasyManua.ls Logo

Agilent Technologies Infinity 1260 - Running a Checkout Sample

Agilent Technologies Infinity 1260
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
72 Agilent 1260 Infinity RID User Manual
4 Using the Refractive Index Detector
Running a Checkout Sample
Running a Checkout Sample
This chapter describes the check out of the Agilent 1260 Infinity Refractive
Index Detector using the Agilent isocratic checkout sample.
1 Turn the detector on.
You are now ready to change the settings of your detector.
2 Set up the instrument with the following chromatographic conditions.
When If you want to checkout the detector
Parts required # p/n Description
1 993967-902 Zorbax Eclipse XDB C18, 150mm x 4.6mm i.d.
1 01080-68704 Agilent isocratic checkout sample
Tab le 16 Chromatographic Conditions
Mobile phases 30% Water, 70% Acetonitrile
Column Zorbax Eclipse XDB C18, 150 mm x 4.6 mm i.d.
Sample Isocratic standard sample
Flow rate 1.5 ml/min
Stroke A 20 µl
Stop time 10 min
Injection volume 20 µl
Column compartment temperature 25 °C
Optical unit Temperature 35 °C
Polarity Positive
Peak Width (Response time) 0.2 min (4 s, standard)

Table of Contents

Related product manuals