EasyManuals Logo

Agilent Technologies Infinity 1260 User Manual

Agilent Technologies Infinity 1260
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #113 background imageLoading...
Page #113 background image
Agilent 1260 Infinity RID User Manual 113
Error Information
7
Refractive Index Detector Specific Error Messages
Lamp Current too Low
Wait Function Timed Out
Wait for temperature or wait for defined signal has not been fulfilled within
the specified time frame.
Probable cause Suggested actions
1 Optical unit cable disconnected.
Please contact your Agilent service
representative.
2 Defective main board.
Please contact your Agilent service
representative.
3 Defective lamp or optics.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Time too short.
Increase time.