Agilent 1260 Infinity RID User Manual 109
Error Information
7
Refractive Index Detector Specific Error Messages
Undecipherable Temperature Signal
Maximum Temperature Exceeded
The maximum heater temperature has been exceeded.
Probable cause Suggested actions
1 Heater cable disconnected.
Please contact your Agilent service
representative.
2 Defective main board.
Please contact your Agilent service
representative.
3 Defective heater.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Defective main board.
Please contact your Agilent service
representative.
2 Defective heater.
Please contact your Agilent service
representative.