Agilent 1260 Infinity RID User Manual 115
Error Information
7
Not-Ready Messages
Unbalanced Diodes
Not Enough Light
Too Much Light
The amount of light falling on the light receiving diodes is too high to generate
a refractive index signal.
Probable cause Suggested actions
1 The diode balance value is outside the
pre-set range -0.5 to + 0.5, an unequal
amount of light is falling on the two light
receiving diodes.
Flush the reference cell with the mobile
phase being used.
Perform the RID Optical Balance procedure
(see “The Optical Balance Procedure” on
page 124).
Probable cause Suggested actions
1 There is insufficient light falling on the light
receiving diodes to generate a refractive
index signal.
Flush the flow cell with the mobile phase being
used to ensure that it is free of air bubbles or
other contamination.
Probable cause Suggested actions
1 The sample cell content is varying too much
from the reference cell.
Purge reference and sample cell.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity 1260 and is the answer not in the manual?

Agilent Technologies Infinity 1260 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity 1260
CategorySecurity Sensors
LanguageEnglish

Related product manuals