EasyManua.ls Logo

Agilent Technologies Infinity 1260 - Refractive Index Calibration

Agilent Technologies Infinity 1260
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
118 Agilent 1260 Infinity RID User Manual
8 Test Functions
Refractive Index Calibration
Refractive Index Calibration
Refractive Index Calibration
The refractive index calibration is based on a sucrose calibration solution,
which has a known refractive index compared to LC grade water. After both
the sample and reference cells have been purged with LC grade water the
sucrose solution is introduced into the flow cell and then the built-in
refractive index calibration functionality is used.
Filling the sample cell with the sucrose calibration solution will give a
theoretical detector response of 512,000 nRIU +/- 5,000 nRIU. The calibration
algorithm will allow the actual detector response, if different, to be changed to
the theoretical value.
The Refractive Index Calibration Procedure
NOTE
Refractive index calibration is only required after exchange of the optical unit or the main
(RIM) - board.
When Recommended after exchange of the optical unit or RIM board.
Tools required Laboratory balance
Parts required # p/n Description
1 DAB/Ph Eur/BP/JP/NF/USP Grade Sucrose
1 9301-1446 Syringe
1 9301-0407 Needle
1 5061-3367 Sample filter
1 0100-1516 Fitting male PEEK, 2/pk

Table of Contents

Related product manuals