16 Agilent 1260 Infinity RID User Manual
1 Introduction to the Refractive Index Detector
Flow Path
Flow Path
The column eluent enters the optical unit through the in port and passes
through a heat exchanger. The combination of the heat exchanger and control
of the optical unit temperature in the range of 5 °C above ambient to 55 °CC
minimizes changes in refractive index due to temperature variations. The
eluent flows through the sample cell and via the same heat exchanger to the
purge valve. With the purge valve in the OFF position the eluent passes to the
recycle valve. If the recycle valve is also in the OFF/WASTE position the eluent
will flow via the waste port into the waste container.
If the recycle valve is in the ON/BOTTLE position the eluent will flow via the
recycle port back to the solvent bottle. The recycle valve can be manually set to
the ON or OFF position or the Automatic recycling after analysis mode can be
enabled. In this mode the recycle valve will automatically switch to the ON
position after each analysis has been completed and return to the OFF
position before the next analysis starts. Using this mode provides the benefits
of uninterrupted flow through the detector without the problems of excessive
solvent usage or the contamination of mobile phase with recycled sample
compounds.
If the purge valve is in the on position the eluent cannot pass immediately to
the recycle valve but will instead flow via a second heat exchanger through the
reference cell and then into the recycle valve (see Figure 5 on page 17).
Periodically switching the purge valve to the on position while only mobile
phase is flowing will ensure that the liquid in the reference cell is as similar as
possible to the flowing solvent. The purge valve can be manually set to the on
position for a defined time or the Automatic purge mode can be enabled. In this
mode the purge valve will automatically switch to the ON position for a
defined purgetime prior to the start of each analysis. If a purgetime is set then a
waittime must also be set to allow the detector baseline to stabilize after the
switching of the purge valve position.
After both the purgetime and waittime have been completed the analysis will
start. If the Automatic zero before analysis mode is enabled the detector output
will be set to zero immediately before the analysis begins.