Agilent 1260 Infinity RID User Manual 111
Error Information
7
Refractive Index Detector Specific Error Messages
Purge Valve Not Connected
When activated no response was received from the purge valve.
Recycle Valve Missing
When activated no response was received from the recycle valve.
Probable cause Suggested actions
1 Purge valve disconnected.
Please contact your Agilent service
representative.
2 Defective purge valve.
Please contact your Agilent service
representative.
3 Defective main board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Recycle valve disconnected.
Please contact your Agilent service
representative.
2 Defective recycle valve.
Please contact your Agilent service
representative.
3 Defective main board.
Please contact your Agilent service
representative.