EasyManua.ls Logo

Agilent Technologies Infinity 1260 - Refractive Index Detector Specific Error Messages

Agilent Technologies Infinity 1260
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 1260 Infinity RID User Manual 107
Error Information
7
Refractive Index Detector Specific Error Messages
Refractive Index Detector Specific Error Messages
Thermal Fuse Open
The thermal fuse of the optical unit heater has failed.
Heater Resistance Too High
The resistance of the heater foil is above the set limit.
Probable cause Suggested actions
1 Heater cable disconnected.
Please contact your Agilent service
representative.
2 Defective main board.
Please contact your Agilent service
representative.
3 Defective thermal fuse.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Heater cable disconnected.
Please contact your Agilent service
representative.
2 Defective main board.
Please contact your Agilent service
representative.
3 Defective heater.
Please contact your Agilent service
representative.

Table of Contents

Related product manuals