Agilent 1260 Infinity RID User Manual 43
Site Requirements and Specifications
2
Physical Specifications
Physical Specifications
Tab le 12 Physical Specifications
Type Specification Comments
Weight 17 kg (38 lbs)
Dimensions (height ×
width × depth)
180 x 345 × 435 mm (7 x 13.5 × 17 inches)
Line voltage 100 – 240 VAC, ± 10% Wide-ranging capability
Line frequency 50 or 60 Hz, ± 5%
Power consumption 160 VA / 65 W / 222 BTU Maximum
Ambient operating
temperature
0–55 °C (32–131 °F)
Ambient non-operating
temperature
-40–70 °C (-4–158 °F)
Humidity < 95%, at 25–40 °C (77–104 °F) Non-condensing
Operating Altitude Up to 2000 m (6562 ft)
Non-operating altitude Up to 4600 m (15091 ft) For storing the module
Safety standards:
IEC, CSA, UL
Installation Category II, Pollution Degree 2 For indoor use only.