108 Agilent 1260 Infinity RID User Manual
7 Error Information
Refractive Index Detector Specific Error Messages
Heater Fuse
The electronic fuse of the heater has been activated.
Wrong Temperature Profile
After turning ON the optical unit heat control, the temperature does not
increase at a sufficiently fast rate to reach the set point.
Probable cause Suggested actions
1 Short in heater circuit.
Power cycle the detector.
2 Defective main board.
Please contact your Agilent service
representative.
3 Defective heater.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Defective main board.
Please contact your Agilent service
representative.
2 Defective heater.
Please contact your Agilent service
representative.