112 Agilent 1260 Infinity RID User Manual
7 Error Information
Refractive Index Detector Specific Error Messages
Lamp Voltage too Low
Lamp Voltage too High
Lamp Current too High
Probable cause Suggested actions
1 Defective main board.
Please contact your Agilent service
representative.
2 Defective lamp or optics.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Contaminated flow cell.
Flush the flow cell.
2 Defective main board.
Please contact your Agilent service
representative.
3 Defective lamp or optics.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Defective main board.
Please contact your Agilent service
representative.
2 Defective lamp or optics.
Please contact your Agilent service
representative.