Agilent 1260 Infinity RID User Manual 123
Test Functions
8
Optical Balance
Optical Balance
Optical Balance
When the sample and reference cells both contain the same liquids an equal
amount of light should fall on each light receiving diode, the diode balance will
equal 0. If this balance of light needs to be corrected the optical balance
procedure can be used.
Diode balance is calculated as follows:
Where:
diode
1
= signal proportional to the amount of light falling on diode
1
diode
2
= signal proportional to the amount of light falling on diode
2
Optical balance adjustment is a manual procedure where the position of the
light beam falling on the light receiving diode is adjusted using the zero glass
adjustment screw.
NOTE
The detector will become not-ready when the diode balance value falls outside the range -
0.5 to + 0.5.
NOTE
Both sample and reference cell must be purged with the same solvent before optical
balance is performed. Prior to performing this procedure, the system must be well
equilibrated.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity 1260 and is the answer not in the manual?

Agilent Technologies Infinity 1260 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity 1260
CategorySecurity Sensors
LanguageEnglish

Related product manuals