EasyManuals Logo

Agilent Technologies Infinity 1260 User Manual

Agilent Technologies Infinity 1260
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #122 background imageLoading...
Page #122 background image
122 Agilent 1260 Infinity RID User Manual
8 Test Functions
Refractive Index Calibration
7 Calibrate refractive index.
a If the detector response differs from the theoretical response of 512,000
nRIU +/- 5,000 nRIU enter the theoretical value (512,000) in the dialog
box. If the detector response is within the theoretical response click OK.
NOTE
Rinse the sample cell with pure water at a minimum of 1.5 mL/min to flush the Sucrose
from the cell and the capillaries. When organic solvent is sequentially applied (without
rinsing), a blockage of capillaries may occur.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity 1260 and is the answer not in the manual?

Agilent Technologies Infinity 1260 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity 1260
CategorySecurity Sensors
LanguageEnglish

Related product manuals