110 Agilent 1260 Infinity RID User Manual
7 Error Information
Refractive Index Detector Specific Error Messages
Purge Valve Fuse Blown
The electronic fuse on the purge valve has been activated.
Recycle Valve Fuse Blown
The electronic fuse on the recycle valve has been activated.
Probable cause Suggested actions
1 Short in purge valve circuit.
Power cycle the module.
2 Defective purge valve.
Please contact your Agilent service
representative.
3 Defective main board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Short in recycle valve circuit.
Power cycle the module.
2 Defective recycle valve.
Please contact your Agilent service
representative.
3 Defective main board.
Please contact your Agilent service
representative.