114 Agilent 1260 Infinity RID User Manual
7 Error Information
Not-Ready Messages
Not-Ready Messages
Not-ready messages are displayed during the wait for a specific condition to be
reached or completed or while a self-test procedure is running. In the event of
such a failure, the yellow status indicator at the front of the detector is
switched ON.
This section describes the meaning of detector not-ready messages.
Purge Time Running
Wait for Purge
Probable cause Suggested actions
1 The purge valve is open, liquid is flowing
through both sample and reference cell.
Allow the reference purge time to elapse.
Probable cause Suggested actions
1 The detector is waiting after the automatic
purge of the reference cell.
Allow the wait time to elapse.