46 Agilent 1260 Infinity RID User Manual
2 Site Requirements and Specifications
Performance Specifications
Environment 0 to 55 °C constant temperature
at < 95% humidity
(non-condensing)
Dimensions 180 mm x 345 mm x 435 mm
(7 x 13.5 x 17 inches)
(height x width x depth)
Weight 17 kg (38 lbs)
Tab le 13 Performance Specifications Agilent 1260 Infinity Refractive Index Detector
Type Specification Comments
NOTE
Based on ASTM method E-1303-95 Practice for Refractive Index Detectors used in Liquid
Chromatography. Reference conditions; optics temperature 35 °C, response time 4 s, flow
1.0 mL/min LC-grade Water, restriction capillary, column compartment temperature 35 °C,
Agilent on-line degasser G1322A, pump and thermostatted column compartment.
Instrument equilibrated for 2 hours.