Agilent 1260 Infinity RID User Manual 67
Using the Refractive Index Detector
4
Operation of the Refractive Index Detector
Analog Output Range: The Analog Output Range group allows you to select the
voltage ranges of the analog output of the refractive index detector. Select
0.1 V to set the full-scale output to 0.1 volts. Select 1 V to set the full-scale
output to 1 volt.
Purge Reference Cell: This parameter is used to exchange the content of the
reference cell in the case of solvent change or reference cell contamination.
Enter a time interval (minutes) to purge the reference cell of the Agilent
1260 Infinity RID. This will be started immediately if you click OK on this
window. Allow additional time for baseline stabilization after purging.
At Power On: When activated, the optical unit heater is turned on
automatically when the RID is switched on. For shortest equilibration
times, Agilent recommends to leave this function always on.
•Automatic Turn On: This function allows you to turn on the optical unit heater
at a specified date and time. It requires that the At Power On function is
turned off. Select Turn Heater on at: to activate the date and time fields, and
enter the date and time in the appropriate fields in the specified format.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity 1260 and is the answer not in the manual?

Agilent Technologies Infinity 1260 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity 1260
CategorySecurity Sensors
LanguageEnglish

Related product manuals