4 Basic Operation
Setting Up an Acquisition Method
79 Agilent Triple Quadrupole LC/MS System User Guide
d Optional. On the Quant tab, mark the Quant Automation check box. The
name of the current Quantitative Analysis method is shown in the list. To
change the Quantitative Analysis method that is connected, click
Browse to select a different method. When the Acquisition method is
saved, the selected Quantitative Analysis method is copied or linked to the
Acquisition method.
e Optional. Mark the Generate Report check box on the Quant tab. Then,
select the Method Path/Template File to use. To print the report, mark the
Print to default printer check box. Mark the Publish check box to create a
CSV file, a TXT file, or a PDF file.
f Optional. For information on the Intelligent Reflex tab, refer to the online
help.
6 Optional. On the Properties tab, set up the properties for this method.
7 Save the method.
a Click Save or Save As.
b If the Audit Trail Settings in Control Panel are set to prompt for a reason
when the method is saved, enter the Reason for creating a new version of
this method. Refer to the MassHunter Control Panel online help for more
information. Click OK.
c If necessary, name the method and click OK.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Triple Quadrupole LC/MS and is the answer not in the manual?

Agilent Technologies Triple Quadrupole LC/MS Specifications

General IconGeneral
BrandAgilent Technologies
ModelTriple Quadrupole LC/MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals