Remote Programming and Monitoring
TM-F1OP-C1XN-01 A-9
Remote Programming and Monitoring
Remote Start/Stop and
Interlock
2.5-15 V signal or TTL-compatible input, selectable
logic. TTL input impedance: 2 k (in series with one
diode drop).
Remote Analog
Programming (Full
Scale Input)
Voltage and current programming inputs (source
must be isolated):
0-5 k, 0-10 k resistances;
0-5 V, 0-10 V (default) voltage sources.
Input impedance (V and I): approximately 100 k.
Remote Monitoring Output voltage and current: 0-5 V, 0-10 V (default);
Output impedance (V and I): approximately 221 Ω.
Output short circuit current: approximately 50 mA.
Remote Programming
and Monitoring
Accuracy
1% zero to full scale output for the default range.
Maximum Remote
Sense Line Drop
Compensation
5 V/line (Line drop is subtracted from total voltage
available at supply output.)
Optional Isolated
Program and
Readback (V&I)
0-5 V (ISOL interface);
input impedance: approx. 200 k;
output impedance: low
Optional Digital
Control
RS-232, GPIB interfaces
TM-F1OP-C1XN-01.book Page 9 Thursday, October 25, 2007 9:04 PM