Appendix A — Test Records A-2 Test Records for Spectrum Analyzer Verification
MS2720T MM PN: 10580-00342 Rev. D A-13
MS2720T Firmware Rev: ______________ Operator: ____________________ Date: ____________
Serial Number: _______________ Options: __________________________________________________
Note
The listed Amplitude Accuracy specifications apply when the instrument is being verified in an
environment between 20°C and 30°C after a 30 minute warm-up period. For an instrument that is not
being verified in a temperature controlled environment, but within a temperature range between
–10°C and 55°C and after a 60 minute warm-up, add ± 1 dB to the Amplitude Accuracy specification
for frequencies from 100 kHz to 40 GHz.
Table A-22. Verification at 8030.1 MHz, Specification = ± 1.3 dB for Opt.713 to 743, ± 1.8 dB for Opt. 709
Input Power Atten Lvl Marker 1 Reading
–30dBm 0dB dBm
–30dBm 5dB dBm
–30 dBm 10 dB dBm
–30 dBm 20 dB dBm
–2 dBm or –5 dBm 30 dB dBm
–2 dBm or –5 dBm 40 dB dBm
–2 dBm or –5 dBm 50 dB dBm
–2 dBm or –5 dBm 60 dB dBm
Table A-23. Verification at 8950.1 MHz, Specification = ± 1.3 dB for Opt.713 to 743, ± 1.8 dB for Opt. 709
Input Power Atten Lvl Marker 1 Reading
–30dBm 0dB dBm
–30dBm 5dB dBm
–30 dBm 10 dB dBm
–30 dBm 20 dB dBm
–2 dBm or –5 dBm 30 dB dBm
–2 dBm or –5 dBm 40 dB dBm
–2 dBm or –5 dBm 50 dB dBm
–2 dBm or –5 dBm 60 dB dBm