EasyManua.ls Logo

Anritsu Spectrum Master MS2720T - Third Order Intercept (TOI); 2400 Mhz TOI Test

Anritsu Spectrum Master MS2720T
174 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 — Spectrum Analyzer Verification 2-11 Third Order Intercept (TOI)
MS2720T MM PN: 10580-00342 Rev. D 2-23
2-11 Third Order Intercept (TOI)
The following test verifies the Third Order Intercept point (also known as TOI or IP3) of the spectrum analyzer
in Spectrum Master Model MS2720T.
2400 MHz TOI Test
Equipment Required for MS2720T with option 709, 713 or 720
Anritsu MG3692x Synthesized Signal Generator (see equipment list for required options)
Frequency Reference Symmetricom Rubisource T&M
Adapter, Anritsu Model 34NN50A
Spectrum Master: MS2720T
Power Meter, Anritsu Model ML2438A
Signal Generator, Agilent Model 8648D
Power Splitter, Aeroflex/Weinschel Model 1870A
Power Sensor, Anritsu Model MA2442D (10 MHz to 18 GHz, –67 dB to +20 dB)
Fixed Attenuator, Aeroflex/Weinschel Model 442
Fixed Attenuator, Aeroflex/Weinschel Model 446
Fixed Attenuator, Aeroflex/Weinschel Model 4420
RF Coaxial Cable, N(m) to N(f), Anritsu Model 15NNF50-1.5C (3)
Equipment Required for MS2720T with option 732 or 743
All of the above, plus
Anritsu 71693-R Ruggedized K(f) to N(f) Adapter

Table of Contents

Other manuals for Anritsu Spectrum Master MS2720T