Chapter 3 — Options Verification 3-5 Option 881 WCDMA/HSDPA Signal Analyzer Verification
MS2720T MM PN: 10580-00342 Rev. D 3-19
WCDMA RF Channel Power Accuracy
The tests in this section can be used to verify the WCDMA RF Channel Power Accuracy in WCDMA/HSDPA
Signal Analyzer Mode.
Procedure
1. Turn on the MG3700A Vector Signal Generator and the ML2438A power meter.
2. Connect the power sensor to the power meter and zero the sensor.
3. Connect the MG3700A, splitter, attenuator, power meter, and sensor as shown in Figure 3-7.
4. Press the On/Off key to turn on the MS2720T and wait until the measurement display appears.
Then press the Shift key, and then press the Mode (9) key to activate the mode selection menu.
5. Use the Up/Down arrow keys to select WCDMA, and press the Enter key.
6. Press the Shift key, the Preset (1) key, and then the Preset submenu key.
7. On the MG3700A, press the Preset key (Yellow key on the upper-left corner).
8. Press the Down Arrow key to select Yes.
9. Press the Set key (Note that two Set keys are available, and they both do the same thing).
10. Press the (F1) submenu key to select “Load File to Memory”.
11. Press the (F1) submenu key again to select “Select Package”.
12. Using the Down Arrow key, step through the selection list until the WCDMA(BS Tx test) option is
highlighted.
13. Press the Set key.
14. Press the (F6) submenu key, Return.
15. Press the Set key.
16. Using the Down Arrow key, step through the selection list until the TestModel_1_16DPCH option is
highlighted.
17. Press the Set key.
18. Press the MOD On/Off
key, and verify that the Modulation indicator on the display is On.
19. Press the Frequency key, then enter 881.5, and press the MHz key.
20. Press the Level key, then enter –2, and press the dBm key.
21. Use the knob to adjust the power meter to read –20 dBm, and record the Power Meter reading in the
“Power Meter Reading” column of Table A-86, “RF Measurements, WCDMA RF Channel Power
Accuracy” on page A-42.
22. On the MS2720T, press the Center Freq submenu key, enter 881.5, and then press the Enter key.
23. Press the Measurements main menu key, then the RF Measurement submenu key, and press ACLR.
24. Press the Amplitude main menu key, then press the Adjust Range submenu key.
25. Record the measured CH 1 power into the “Measured RF Channel Power” column of Table A-86.
26. Calculate the RF Channel Power Error using the following formula:
RF Channel Power Error (dB) = Measured RF Channel Power – 0.246 – Power Meter reading
27. Record calculated value to the “RF Channel Power Error” column of Table A-86, and verify that it is
within specifications (± 1.25 dB).
28. Repeat Step 19 through Step 27 for the other frequencies.

Table of Contents

Other manuals for Anritsu Spectrum Master MS2720T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu Spectrum Master MS2720T and is the answer not in the manual?

Anritsu Spectrum Master MS2720T Specifications

General IconGeneral
BrandAnritsu
ModelSpectrum Master MS2720T
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals