EasyManuals Logo

Anritsu Spectrum Master MS2720T Maintenance Manual

Anritsu Spectrum Master MS2720T
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #22 background imageLoading...
Page #22 background image
2-5 Spurious Response Verification Chapter 2 — Spectrum Analyzer Verification
2-8 PN: 10580-00342 Rev. D MS2720T MM
22. Press the Marker main menu key and select the Peak Search submenu key.
23. Record the amplitude for 50.1 MHz. Use Table A-3, “Spectrum Analyzer Second Harmonic Distortion”
on page A-2.
24. Press the Freq main menu key and select the Center Freq submenu key.
25. Use the keypad to enter 100.2 and press the MHz submenu key.
26. Press the Shift key and then press the Trace (5) key, then select the Trace A Operations submenu key.
27. Press the # of Averages submenu key, use the keypad to enter 5 and then press the Enter key.
28. Wait until the Trace Count displays 5/5.
29. Press the Marker main menu key and select the Peak Search submenu key.
30. Record the amplitude for 100.2 MHz in the test records. Use Table A-3.
31. Calculate the second Harmonic level in dBc by subtracting the 50.1 MHz amplitude from the 100.2 MHz
amplitude using this formula:
Second Harmonic Level Amplitude @ 100.2 MHz = 100.2 MHz amplitude – 50.1 MHz amplitude = ______ dBc
32. Verify that the calculated Second Harmonic Level is within the specs listed below for the applicable
model under test and record in the test records. Use Table A-3.
For MS2720T –54 dBc

Table of Contents

Other manuals for Anritsu Spectrum Master MS2720T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu Spectrum Master MS2720T and is the answer not in the manual?

Anritsu Spectrum Master MS2720T Specifications

General IconGeneral
BrandAnritsu
ModelSpectrum Master MS2720T
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals