Spectrum Analyzer 2-22 Spurious Emissions (Out-of-Band)
Spectrum Analyzer MG PN: 10580-00349 Rev. H 2-33
6. Press the Stop Freq submenu key and enter a value for the end point of the segment.
7. Return to the Spurious Emissions submenu by pressing the touch screen shortcut in the
upper right corner of the display, where the words Spectrum Analyzer are located.
You can also select the Spurious Emissions submenu by pressing the Shift key followed
by the Measure (4) key, then pressing the Masks and C/I submenu key followed by the
Spurious Emissions (Out-of-Band) key.
8. Press Segment Setup.
9. Press Segment Start Amplitude and enter the desired amplitude for the first point of the
segment.
10. Press Segment Stop Amplitude and enter the desired amplitude for the last point of the
segment.
11. Press Save Current Settings to Selected Segment.
12. Repeat Step 4 through Step 11 as necessary to add more segments.
13. Press Save Mask and enter the mask name in the Filename entry field. Use the existing
mask name if you wish to overwrite the file. If needed, select Spurious Emission Mask
from the Filetype pull-down menu. Press Enter.
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com

Table of Contents

Other manuals for Anritsu Spectrum Master MS2720T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu Spectrum Master MS2720T and is the answer not in the manual?

Anritsu Spectrum Master MS2720T Specifications

General IconGeneral
BrandAnritsu
ModelSpectrum Master MS2720T
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals