Specifications A-3
Sample Temperature Information
Temperature
The table below lists sample temperature information.
Note Sample temperatures are displayed in degrees Celsius to the nearest 0.1 °C.
Printer Specifications
Serial Interface
Board Specifications
The 2720 Thermal Cycler can use any printer with a serial (RS-232C) interface board
with the following parameters.
Cable Part Number
Connect the printer to the instrument port with printer cable part number N805-1326.
Sample Temperature Range 4.0 to 99.9 °C.
Temperature Calibration Traceable to National Institute of Standards and
Technology (NIST).
Baud Rate 9600
Parity NONE
Data Bits 8
Stop Bits 1