2-8 Getting Started
Starting and Stopping a Run
Before You Begin
Let’s assume that we’ve loaded our samples properly and we’re ready to start a run.
Procedure
To start, then stop a run :
Step Action
1 From the Main menu press F1 (Run).
The Stored Methods screen appears.
Since we have only one method (“test01”), it’s already highlighted (selected).
2 Press F1 (Start).
The Select Method Options screen appears.
3 Change the reaction volume to 40, then press F1 (Start).
The Run Time screen appears.
Methods
test01
User
adrian
F1 F3 F4 F5F2
Start CancelSortUserView
Size
11
LastUsed
01/04/00
Select Method Options
Enter a value from 5 to 100 µL
Reaction Volume: 50 µL
F1 F3 F4 F5F2
Start Cancel
Sample temperature increments
Flashes to warn you
Sample 56.7 Cycle 1 of 25 HOT
55.0
94.0
2:00
10:00
55.0
0:30
72.0
0:30
72.0
5:00
4.0
94.0
0:30
F1 F3 F4 F5F2
Pause Info