EasyManua.ls Logo

Applied Biosystems 2720 - Starting a Run

Applied Biosystems 2720
122 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Runs 3-7
Starting a Run
Procedure
To start a run:
Step Action
1 Load your samples, as described in “Loading Samples” on page 3-5.
Note Disposables must be used.
2 If the instrument power is not on, press the power-on button at the back of the
instrument.
3 From the Main Menu press F1 (Run).
The Stored Methods screen appears.
4 Select a method by highlighting one of the displayed methods.
If the method you want is not displayed, see “Handling Methods” on page 4-18.
5 Press F1 (Start).
The Select Method Options screen appears.
6 Enter the reaction volume, then press F1 (Start).
If the temperature of the heated cover is less than 103 °C, the screen shown below
appears.
Methods
appl01
User
adrian
F1 F3 F4 F5F2
Start CancelSortUserView
Size
11
LastUsed
03/04/00
exp000
exp001
exp002
adrian
adrian
adrian
10
12
13
02/22/00
02/10/00
02/02/00
Select Method Options
Enter a value from 5 to 100 µL
Reaction Volume: 50 µL
F1 F3 F4 F5F2
Start Cancel
Current temperature: 65°C
The run will begin when the
heated cover reaches 103°C.
Please wait. Cover is heating
F1 F3 F4 F5F2
Cancel

Table of Contents

Related product manuals