Troubleshooting 7-3
Display Screen Error Messages
User Input Error
Messages
The following table lists user input error messages, a description of the message, and
recommended action.
Message Description Recommended Action
Can only enter an infinity hold
at end
A method can only have a Hold segment
with an infinity hold as the last segment in a
method.
This message occurs when you try to enter
an infinity hold segment in the middle of a
method.
Assign finite time segments to holds
within a method.
Delete your methods first User tried to delete a user name that has
methods stored under it.
Delete or transfer the associated
methods before deleting a user name.
Enter a name or CANCEL You did not enter at least one character on
the User Name screen before pressing the
Accept key.
Enter the user name to which the
desired method is assigned.
Enter oligo sequence Incomplete TmCalc data. Enter a value in the P1 and P2 fields of
the TmCalc.
Enter user and method names
or CANCEL
You did not enter a user name and a method
name before storing a method.
Specify the method name and choose
a user to store a method.
Field is full You tried to enter more data in a field than
the field size allows.
Reenter data within the specifications
of the field.
Infinity hold not allowed in
cycle
A method can have an infinity hold segment
as the last segment in the method.
This message occurs when you tried to
enter an infinity time in a cycle segment.
Use finite values for cycle segments
within the method.
Invalid password/pin# You entered an incorrect PIN#. Enter the correct PIN#.
List of user names is full The maximum number of users has been
entered into the system.
Delete unused user names.
Maximum of 6 segments
allowed
You tried to insert more than six temperature
control parameters into a hold or cycle.
Do not assign more than six hold or
cycle parameters to a method.
Method battery RAM initialized Stored methods have been reset due to
hardware or software failure.
Not all methods may be lost.
Check method directory.
Refer to our Web site.
a
Method requires at least one
segment
You deleted all temperature control
parameters in a method.
A method must have at least one time and
temperature parameter.
Review and correct the method to
include the temperature parameter(s).
No seconds in time field You did not include seconds in the time field. Include seconds when entering the
time.
Not enough method memory
left
This error occurs:
When you attempt to exceed the limit of
137 methods.
When you attempt to store or create a
new method that is larger than the
available storage space.
Determine how much storage
memory is available on the
instrument.
Delete or store rarely used methods
elsewhere.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 2720 and is the answer not in the manual?

Applied Biosystems 2720 Specifications

General IconGeneral
BrandApplied Biosystems
Model2720
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals