7-4 Troubleshooting
Not implemented yet The feature is not implemented in the
current firmware version.
Upgrade firmware when the new
version is available from Applied
Biosystems.
Printer not responding The printer has been disconnected or is off
line.
Check printer connections and power
switch.
Remove infinity hold first A method can have an infinity hold segment
as the last segment in the method.
This message occurs when a user tries to
add a segment after one which contains an
infinity hold.
Add segments prior to the post-PCR
infinity hold.
T
m
temperature out of range T
m
out of range. Check input value and retry.
User name already defined You entered a user name that already exists. Do not duplicate user names.
Valid range is You entered a number out of range.
The message include the valid range limits.
Re-enter a value within the parameters
of the field.