Utilities 5-9
Running the Calibration Verification Test
Why Use This Test?
Use this test to verify the temperature calibration of your 2720 Thermal Cycler.
Equipment Required
This test requires the 0.2-mL Temperature Verification System (P/N 4317939) for
96-well thermal cyclers.
For more information on performing the test, refer to the instructions included with
your Temperature Verification System. Follow the procedure for the GeneAmp
®
PCR
System 9700.
Running the Temperature Non-Uniformity Test
Why Use This Test?
Use this test to verify the temperature non-uniformity of the sample block in the 2720
Thermal Cycler.
Equipment Required
This test requires the 0.2-mL Temperature Verification System (P/N 4317939) for
96-well thermal cyclers.
For more information on performing the test, refer to the instructions included with
your Temperature Verification System. Follow the procedure for the GeneAmp
®
PCR
System 9700.