3-8 Runs
7 Wait for the heated cover to reach 103 °C.
The Run Time screen displays and the method you selected starts running.
8 Optional. To find out when the run will end, press F4 (Info).
The Method Information screen appears.
When you are through viewing this screen, press F4 (Return) to redisplay the Run
Time screen.
To start a run: (continued)
Step Action
Sample 56.7 Cycle 1 of 25 HOT
55.0
94.0
2:00
10:00
55.0
0:30
72.0
0:30
72.0
5:00
4.0
94.0
0:30
F1 F3 F4 F5F2
Pause Info
02:32 PM Information 55.2°C
User: lisa
Run started at 01:32:30 PM, 03/01/00.
Run will end at 06:35:30 PM, 03/01/00.
Reaction vol: 50 µL
Method: General PCR