3-12 Runs
History File Records
The following table lists the history file record, a description of the record, and the type
of record.
Record Description Type
History of method xxxxxxxxxxxxxxx
User xxxxxx Reaction volume xxx µL
Run started at hh:mm:ss am mm/dd/yy
Run ended at hh:mm:ss am mm/dd/yy
Length of run hh:mm:ss RampSpeed: 9600
This header record is always created. Report
Power failure in Cycle xx at Setpt xx.
Power failed at hh:mm:ss am for hh:mm:ss.
Run resumed at hh:mm:ss am
There was a power failure during a
specified point in a cycle.
The message, for >18, indicates that
the power was off for more than 18
hours.
Report
Drift error in Cycle xx Setpt xx Repxx. Temperature
drifted x.x°c from setpt
Block drift error.
The block has drifted ± 2 °C from set
point during the hold segment of a run.
Report
Cover error in Cycle xx Setpt xx Repxx.
Heated cover at xx.x°c
Heated cover drift error.
The cover has drifted ± 5 °C from
105 °C anytime during the run.
Report
Sensor error in Cycle xx Setpt xx Repxx.
Block sensor failure.
Block sensor failure. Fatal error.
Access our Web
site.
a
Sensor error in Cycle xx Setpt xx Repxx.
Cover sensor failure.
Heated cover sensor failure. Fatal error.
Access our Web
site.
a
Setpt error in Cycle xx Setpt xx Repxx.
Could not reach xx.x in hh:mm:ss
This setpoint error is only logged for
setpoints above 15 °C.
The limit is 5 times the normal ramping
time.
Fatal error.
Access our Web
site.
a
Program pause in Cycle xx Setpt xx Rep xxx Method
paused at xx °C for hh:mm:ss
A programmed pause was
encountered.
Report
Manual pause in Cycle xx Setpt xx Rep xxx Method
paused at xx °C for hh:mm:ss
You paused the run. Report
a. http://www.appliedbiosystems.com/support/contact