5-12 Utilities
3 At the conclusion of the test, the display indicates test results and whether the test
results passed or failed.
The following table lists the passing ranges for the Cycle Test.
To run the Cycle Test: (continued)
Step Action
Cycle Performance Pass
Average Cycle Time: xxx.x sec
F1 F3 F4 F5F2
CancelPrint
Cycle Time STD:
x.x sec
Average Cycle Time < 129 sec
Cycle Time STD <
1.0 sec