Appendix A
Recommended Maintenance Schedule
106 Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Notes
Semi-Annual
Maintenance
Tasks
Perform a background calibration (see page 49)
Perform an optical calibration on 7500/7500 Fast systems (see page 59)
Perform a region of interest (ROI) calibration (see page 35)
Perform a pure dye calibration (see page 67)
Note: You can run a background calibration to check for contamination.If any parts of
the optics are replaced or moved, you must run a background calibration, an ROI
calibration, an optical calibration (7500/7500 Fast only), and a pure dye calibration.
Miscellaneous
Maintenance
Tasks
Perform the following tasks as needed to resolve problems as they arise:
Decontaminate the sample block (see page 108)
Move the 7300/7500/7500 Fast system (see page 115)
Replace the halogen bulb (see page 117)
Replace the instrument fuses (see page 122)
Update the Microsoft Windows operating system (see page 124)
6 Months
January
Su M T W Th F S Su M T W Th F S
July

Table of Contents

Other manuals for Applied Biosystems 7500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7500 and is the answer not in the manual?

Applied Biosystems 7500 Specifications

General IconGeneral
Detection MethodFluorescence
Sample Capacity96-well plate
Thermal CyclerYes
Temperature Uniformity±0.5°C
TypeReal-Time PCR System
Thermal Accuracy±0.25°C
Detection Channels5
Weight45 kg
Power Requirements100-240 V, 50/60 Hz
Dynamic Range9 logs
Reaction Volume Range10-100 µL
Temperature Range4°C to 100°C
Thermal Cycler Block96-well
SoftwareSDS Software
DetectionMultiplex (up to 5 colors)

Related product manuals