EasyManuals Logo

Applied Biosystems 7500 User Manual

Applied Biosystems 7500
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #58 background imageLoading...
Page #58 background image
Chapter 3 Installing the Software and Powering On the Instrument
Troubleshooting
32 Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Notes
Troubleshooting
This section contains:
Troubleshooting – Front Panel Indicators page 32
Troubleshooting – Function Tests page 33
Troubleshooting – Front Panel Indicators
Troubleshooting – Front Panel Indicators
Condition: The red Error indicator is lit.
1. Press on the instrument door to ensure that it is closed.
If the green Power On indicator lights up, the open instrument
door caused the error, and installation continues.
2. If the red Error indicator remains lit:
a. Open the instrument door.
b. Pull the heated cover door to verify that it is closed.
c. Close the instrument door.
If the green Power On indicator lights up, the open heated
cover door caused the error, and installation continues.
3. If the red Error indicator remains lit, verify that the Windows
desktop is displayed on the computer. If the Windows desktop
is not displayed:
a. Power off the 7300/7500/7500 Fast instrument.
b. Restart the computer.
c. Wait until the Windows desktop appears.
d. Power on the 7300/7500/7500 Fast instrument.
If the green Power On indicator lights up, installation
continues.
Instrument door
Heated cover door
Windows desktop

Table of Contents

Other manuals for Applied Biosystems 7500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7500 and is the answer not in the manual?

Applied Biosystems 7500 Specifications

General IconGeneral
Detection MethodFluorescence
Sample Capacity96-well plate
Thermal CyclerYes
Temperature Uniformity±0.5°C
TypeReal-Time PCR System
Thermal Accuracy±0.25°C
Detection Channels5
Weight45 kg
Power Requirements100-240 V, 50/60 Hz
Dynamic Range9 logs
Reaction Volume Range10-100 µL
Temperature Range4°C to 100°C
Thermal Cycler Block96-well
SoftwareSDS Software
DetectionMultiplex (up to 5 colors)

Related product manuals