Appendix C
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide 127
Notes
C
Creating a Custom Pure Dye Plate
Overview
The Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System can be used to
run assays designed with custom dyes (dyes not manufactured by Applied Biosystems).
Custom dyes must fluoresce within the spectral range measured by the instrument:
500 to 650 nm for 7300 systems
500 to 700 nm for the 7500/7500 Fast systems
Before using custom dyes with the 7300/7500/7500 Fast instrument, you must:
Determine optimum dye concentration
Create a custom pure dye plate
Add the custom dye to the software
Perform a pure dye calibration (see Chapter 6, “Performing the Pure Dye
Calibration.”
Materials
Required
Optical Adhesive Cover
or Optical Flat Caps
ABI PRISM
®
Optical
96-Well Reaction Plate
and Optical 96-Well Fast
Plate
Custom dye(s)
Powder-free
gloves
Safety
glasses
Tubes
(2-mL and 10-mL)
Pipettors and pipette tips
(200-µL and 1000-µL) Water,
deionized
GR2170
Centrifuge with plate
adapter

Table of Contents

Other manuals for Applied Biosystems 7500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7500 and is the answer not in the manual?

Applied Biosystems 7500 Specifications

General IconGeneral
Detection MethodFluorescence
Sample Capacity96-well plate
Thermal CyclerYes
Temperature Uniformity±0.5°C
TypeReal-Time PCR System
Thermal Accuracy±0.25°C
Detection Channels5
Weight45 kg
Power Requirements100-240 V, 50/60 Hz
Dynamic Range9 logs
Reaction Volume Range10-100 µL
Temperature Range4°C to 100°C
Thermal Cycler Block96-well
SoftwareSDS Software
DetectionMultiplex (up to 5 colors)

Related product manuals