EasyManuals Logo

Applied Biosystems 7500 User Manual

Applied Biosystems 7500
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #143 background imageLoading...
Page #143 background image
Appendix A
Monitoring Lamp Status
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide 117
Notes
A
Monitoring Lamp Status
Checking Status To determine whether the halogen lamp has enough electrical current, select
InstrumentLamp Status/Replacement.
In the
Lamp
Status/Replacement dialog
box, the Lamp Current: field
indicates a percentage figure
for the electrical current.The
Condition: field indicates one
of the following:
Good The lamp is
functioning well. There
is no need to replace the
lamp bulb at this time.
Click .
Failed The lamp bulb must be replaced. Click , then replace the lamp as
explained below.
Change Soon The lamp bulb usage is above 2000 hours. It is recommended to
replace the lamp soon. Click , then decide whether to replace the lamp.
If the bulb needs replacement, see “Replacing the Halogen Lamp” on page 118.
Warning
Messages
Three warning messages can be displayed before or during a run that indicate low lamp
current:
Message Description
Warning – Cannot detect
sufficient current from lamp.
Either lamp is not installed
properly or needs to be
replaced.
Displayed at the start of a run if the lamp current has fallen
below the acceptable percentage.
You cannot proceed with the run. You must replace the
halogen bulb.
Warning – Cannot detect
sufficient current from lamp.
Either lamp is not installed
properly or needs to be
replaced.
Displayed if the lamp current falls below the acceptable
percentage during a run. The run is terminated.
Click OK in the message box, inspect the Instrument Log,
then replace the lamp bulb.
You cannot proceed with the run. You must replace the
halogen bulb.
Warning - The lamp usage has
exceeded 2000 hours. We
recommend replacing the lamp
soon to ensure optimal assay
performance.
Displayed at the start of a run if the lamp usage exceeds
2000 hours.
Click Cancel Run, then replace the lamp, or click Continue
Run.

Table of Contents

Other manuals for Applied Biosystems 7500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7500 and is the answer not in the manual?

Applied Biosystems 7500 Specifications

General IconGeneral
Detection MethodFluorescence
Sample Capacity96-well plate
Thermal CyclerYes
Temperature Uniformity±0.5°C
TypeReal-Time PCR System
Thermal Accuracy±0.25°C
Detection Channels5
Weight45 kg
Power Requirements100-240 V, 50/60 Hz
Dynamic Range9 logs
Reaction Volume Range10-100 µL
Temperature Range4°C to 100°C
Thermal Cycler Block96-well
SoftwareSDS Software
DetectionMultiplex (up to 5 colors)

Related product manuals