Chapter 8 Powering Off the 7300/7500/7500 Fast System
Performing a Long-Term Shutdown
104 Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Notes
Performing a Long-Term Shutdown
Perform the long-term shutdown procedure if the
instrument will be inactive for more than 7 days.
Time Required
5 minutes
Materials Required
Plate, packaging (from step 8 on page 21)
Performing a Long-term Shutdown
1. Press the tray to open it.
2. If the tray contains a plate, remove it.
3. Load the packaging plate into the tray.
Note: If the shipping plate is not available,
substitute an unused reaction plate. During
storage, the instrument optics block rests on the
plate to protect the optics block.
4. Press the tray to move it into the instrument.
5. Press the instrument power button.
6. Power off the computer and monitor:
a.
Select
Shut Down
.
b. In the Shut Down Windows dialog box (not
shown), select
Shut Down
.
c. Power off the monitor.
Su M T W Th F S
> 1 week
Load

Table of Contents

Other manuals for Applied Biosystems 7500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7500 and is the answer not in the manual?

Applied Biosystems 7500 Specifications

General IconGeneral
Detection MethodFluorescence
Sample Capacity96-well plate
Thermal CyclerYes
Temperature Uniformity±0.5°C
TypeReal-Time PCR System
Thermal Accuracy±0.25°C
Detection Channels5
Weight45 kg
Power Requirements100-240 V, 50/60 Hz
Dynamic Range9 logs
Reaction Volume Range10-100 µL
Temperature Range4°C to 100°C
Thermal Cycler Block96-well
SoftwareSDS Software
DetectionMultiplex (up to 5 colors)